Análise química laboratorial da superfície de implantes após contato com pinça de titânio através do EDS estudo-piloto / Chemical analysis (EDS) of dental implant surface after contact with titanium tweezers a pilot study
ImplantNewsPerio
; 2(6): 1050-1059, nov.-dez. 2017. ilus, tab
Article
en Pt
| LILACS, BBO
| ID: biblio-880834
Biblioteca responsable:
BR510.1
RESUMO
Objetivo:
analisar a composição química da superfície de três implantes antes e depois de manter contato com pinça de titânio, e verificar se níveis importantes de elementos químicos contaminantes foram adsorvidos aos implantes. Material emétodos:
através de MEV (microscópio eletrônico de varredura) e EDS (espectrometria de energia dispersiva), as superfícies dos implantes foram caracterizadas quimicamente antes e após o contato com pinça de titânio utilizada rotineiramente em uma clínica odontológica.Resultados:
houve presença de titânio com 99,51% de média nas superfícies antes do contato com pinça de titânio, e 99,43% de média de titânio após o contato com pinça de titânio. Elementos como ferro, silício, cloro e alumínio foram encontrados em concentrações menores do que 0,55%, tanto antes quanto depois da manipulação com a pinça.Conclusão:
não houve contaminação química detectável pela metodologia empregada neste estudo da superfície dos implantes após o contato com a pinça de titânio, porém, uma amostra mais significativa e o emprego de critérios que possam definir com mais precisão as áreas de análise parecem ser necessários para resultados ainda mais consistentes.ABSTRACT
Objective:
to analyze the chemical composition surface of three implants before and after titanium tweezers contact and check whether important levels of contaminants chemical elements were adsorbed to the implants. Material andmethods:
through SEM (scanning electron microscope) and EDS (energy dispersive spectroscopy) the implants surfaces were characterized chemically before and after contact with titanium tweezers frequently used in dental clinic.Results:
the surfaces before contact with the titanium tweezers showed the presence of titanium 99.51% average and 99.43% average of titanium after contact with the titanium clamp. Elements such as iron, silicon, chlorine and aluminum have been found in lower concentrations than 0.55%, both before and after handling with forceps.Conclusion:
therefore, no chemical contamination was detected by the method used in this study of the implant surface after contact with titanium tweezers, however, a more signifi cant sample and the use of criteria that can defi ne more precisely the areas of analysis, appear to be required for further consistent results.Palabras clave
Buscar en Google
Colección:
01-internacional
Base de datos:
BBO
/
LILACS
Asunto principal:
Titanio
/
Técnicas de Química Analítica
/
Implantación Dental
/
Contaminación Ambiental
/
Fenómenos Químicos
Tipo de estudio:
Evaluation_studies
Aspecto:
Determinantes_sociais_saude
Límite:
Female
/
Humans
/
Male
Idioma:
Pt
Revista:
ImplantNewsPerio
Asunto de la revista:
Odontologia
Año:
2017
Tipo del documento:
Article
País de afiliación:
Brasil
Pais de publicación:
Brasil