Talanta ; 116: 1026-32, 2013 Nov 15.
Artigo em Inglês | MEDLINE | ID: mdl-24148511

RESUMO

In this paper, we present two new electrochemical methods for fast and simultaneous determination of codeine (CO) and diclofenac (DCF). The first one is based on batch injection analysis with amperometric detection (BIA-MPA) and the second one is based on capillary electrophoresis with capacitively coupled contactless conductivity detection (CE-C(4)D). The proposed BIA-MPA method is highly-precise (RSD of 1.1% and 0.9% for DCF and CO, respectively; n=10), fast (300 injections h(-1)) and has low detection limits (1.1 and 1.0 µmol L(-1) for DCF and CO, respectively). The proposed CE-C(4)D method allows the determination of CO and DFC in less than 1 min with high precision (RSD of 0.3% and 0.7% for DCF and CO, respectively; n=10) and low detection limits (11 and 21 µmol L(-1) for DCF and CO, respectively). Both proposed methods were applied to the determination of CO and DCF in pharmaceutical samples with similar results to those achieved by high-performance liquid chromatography (HPLC) at a 95% confidence level.


Assuntos
Codeína/análise , Diclofenaco/análise , Técnicas Eletroquímicas/instrumentação , Eletroforese Capilar/instrumentação , Comprimidos/química , Cromatografia Líquida de Alta Pressão , Condutividade Elétrica , Análise de Injeção de Fluxo , Limite de Detecção , Fatores de Tempo