Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 5 de 5
Filtrar
Mais filtros











Base de dados
Intervalo de ano de publicação
1.
J Endod ; 18(7): 336-9, 1992 Jul.
Artigo em Inglês | MEDLINE | ID: mdl-1402596

RESUMO

Canal Master instruments have a short, fluted cutting area and a small flexible shaft; they may be predisposed to rapid wear and breakage. This study examined new and used files as to tip, flute, and shaft design when new and with increasing time of canal preparation. One hundred files prepared 140 curved canals. Sizes #20, #40, and #60 files were examined unused and after 1, 3, 5, and 7 min of use. Evaluation was for uniformity when new and for deterioration and breakage with usage. The smaller sizes (#20 and #40) had some inconsistencies when new and were most predisposed to wear and breakage with time. Findings suggest that smaller files could have had improved quality control by the manufacturer. Also, they should be used with caution and discarded after short times of use in small, curved canals. This should minimize instrument separation and maximize cutting efficiency.


Assuntos
Preparo da Cavidade Dentária/instrumentação , Tratamento do Canal Radicular/instrumentação , Falha de Equipamento , Microscopia Eletrônica de Varredura , Controle de Qualidade , Propriedades de Superfície
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA