1.
Appl Radiat Isot ; 68(4-5): 954-6, 2010.
Artigo em Inglês | MEDLINE | ID: mdl-20093037

RESUMO

Thermoluminescent response of LiF:Mg,Cu,P+PTFE under clinical photon irradiation was obtained. Thermoluminescent dosimeters (TLDs) were irradiated for determining entrance surface dose (ESD) in a solid water phantom when using standard clinical adult treatment protocols. A Monte Carlo simulation of photon interaction with matter was performed and absorbed dose determined. ESD calculated by MCNPX code was greater than those determined by direct measurements in phantom. The results obtained open the possibility for using this material as a TLDs in medical accelerators.


Assuntos
Modelos Estatísticos , Método de Monte Carlo , Dosimetria Termoluminescente/instrumentação , Simulação por Computador , Fótons/uso terapêutico , Prótons , Reprodutibilidade dos Testes , Sensibilidade e Especificidade