1.
J Agric Food Chem ; 51(24): 6945-8, 2003 Nov 19.
Artigo em Inglês | MEDLINE | ID: mdl-14611151

RESUMO

This work refers to a very easy to implementate flow injection system with potentiometric detection for l-glutamate determination in food samples. The proposed procedure is based on measurement of carbon dioxide produced by decarboxylation of l-glutamate catalyzed by l-glutamate decarboxylase (E.C. 4.1.1.1.5) from Cucurbita maxima (pumpkin). The FI potentiometric system includes an enzymatic reactor with a length of 8 cm and thickness of 5 mm packed with 200 mg of a C. maxima outer layer cut in to small pieces. The proposed procedure allowed l-glutamate determinations in the concentration interval of 10-100 mmol L(-1) for an injected sample volume of 50 microL. A phosphate buffer (0.1 mol L(-1), pH 5.5) solution flowing at 1.4 mL min(-1) was used as the carrier solution in the system. The results obtained in the analysis of food samples revealed a relative error lower than 5% when compared with those provided by the spectrophometric reference procedure. The immobilized reactor retained its initial activity for 21 days. It was possible to measure 40 samples/h with the flow system proposed.


Assuntos
Cucurbita/enzimologia , Análise de Injeção de Fluxo/métodos , Análise de Alimentos/métodos , Glutamato Descarboxilase , Ácido Glutâmico/análise , Análise de Injeção de Fluxo/instrumentação , Glutamato Descarboxilase/metabolismo , Indicadores e Reagentes , Cinética , Potenciometria
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA