1.
J Appl Crystallogr ; 56(Pt 5): 1480-1484, 2023 Oct 01.
Artículo en Inglés | MEDLINE | ID: mdl-37791352

RESUMEN

The general-purpose powder diffractometer beamline (BL2-1) at the Stanford Synchrotron Radiation Lightsource (SSRL) is described. The evolution of design and performance of BL2-1 are presented, in addition to current operating specifications, applications and measurement capabilities. Recent developments involve a robotic sample changer enabling high-throughput X-ray diffraction measurements, applicable to mail-in and remote operations. In situ and operando capabilities to measure samples with different form factors (e.g. capillary, flat plate or thin film, and transmission) and under variable experimental conditions are discussed. Several example datasets and accompanying Rietveld refinements are presented.

2.
J Appl Crystallogr ; 54(Pt 1): 132-141, 2021 Feb 01.
Artículo en Inglés | MEDLINE | ID: mdl-33833644

RESUMEN

An automatic sample changer system for measurements of large numbers of liquid samples - the µDrop Sample Changer - is presented. It is based on a robotic arm equipped with a pipetting mechanism, which is combined with a novel drop-based sample holder. In this holder a drop of liquid is suspended between two parallel plates by surface tension. The absence of a transfer line benefits the cleaning, improving the background as well as making it faster and more efficient than most comparable capillary-based systems. The µDrop Sample Changer reaches cycle times below 35 s and can process up to 480 samples in a single run. Sample handling is very reliable, with a drop misplacement chance of about 0.2%. Very low sample volumes (<20 µl) are needed and repeatable measurements were performed down to 6 µl. Using measurements of bovine serum albumin and lysozyme, the performance of the instrument and quality of the gathered data of low and high concentrations of proteins are presented. The temperature of samples can also be controlled during storage and during measurement, which is demonstrated by observing a phase transition of a mesophase-forming lipid solution. The instrument has been developed for use in small-angle X-ray scattering experiments, which is a well established technique for measuring (macro-)molecules. It is commonly used in biological studies, where often large sets of rare samples have to be measured.

3.
J Synchrotron Radiat ; 27(Pt 3): 860-863, 2020 May 01.
Artículo en Inglés | MEDLINE | ID: mdl-32381791

RESUMEN

In this paper, the design and functionalities of the high-throughput TELL sample exchange system for macromolecular crystallography is presented. TELL was developed at the Paul Scherrer Institute with a focus on speed, storage capacity and reliability to serve the three macromolecular crystallography beamlines of the Swiss Light Source, as well as the SwissMX instrument at SwissFEL.


Asunto(s)
Cristalografía por Rayos X/instrumentación , Sustancias Macromoleculares/química , Diseño de Equipo , Reproducibilidad de los Resultados , Robótica/instrumentación , Sincrotrones/instrumentación
4.
J Radioanal Nucl Chem ; 309: 27-38, 2016.
Artículo en Inglés | MEDLINE | ID: mdl-27375310

RESUMEN

The present status of development of software packages and equipment designed for automation of NAA at the reactor IBR-2 of FLNP, JINR, Dubna, RF, is described. The NAA database, construction of sample changers and software for automation of spectra measurement and calculation of concentrations are presented. Automation of QC procedures is integrated in the software developed. Details of the design are shown.