1.
Diabetes Technol Ther ; 25(8): 554-558, 2023 08.
Artículo en Inglés | MEDLINE | ID: mdl-37184526

RESUMEN

The aim of this study was to observe the 1-year clinical outcomes of people with type 1 diabetes who switched from multiple daily injection + blood glucose meter to an advanced hybrid closed-loop (AHCL) system (Medtronic MiniMed™ 780G system [MM 780G]). In addition, the effect of changing at month 6 to a calibration-free sensor (Guardian™ 4 Sensor [G4S]) was evaluated. Eighteen participants (10 men, age 40.9 ± 7.6 years) completed 1 year of MM 780G use. Time in range (TIR; 70-180 mg/dL) remained stable and ranged from 83.2% in month 9 to 84.8% in month 3. There was no difference between TIR at 3 months before switching versus 3 months after switching to G4S (P = 0.614). AHCL system in adults significantly improves glycemic outcomes. This improved glycemic control was maintained over the 12 months. Switching to a calibration-free sensor (G4S) did not affect outcomes but required less patient involvement.


Asunto(s)
Diabetes Mellitus Tipo 1 , Masculino , Adulto , Humanos , Persona de Mediana Edad , Diabetes Mellitus Tipo 1/tratamiento farmacológico , Hipoglucemiantes/uso terapéutico , Glucemia , Insulina/uso terapéutico , Estudios de Seguimiento , Sistemas de Infusión de Insulina , Automonitorización de la Glucosa Sanguínea