1.
Artículo en Inglés | MEDLINE | ID: mdl-17709069

RESUMEN

OBJECTIVE: The aim of this in vitro study was to compare the capacity of the Root ZX-II and Mini Apex Locator, electronic apex locators (EALS), to prevent overestimated working length. STUDY DESIGN: Forty extracted human teeth were used for the study. The cervical portion of each canal was flared using Gates Glidden drills and the teeth were embedded in an alginate model. Canals were irrigated with 2.5% sodium hypochlorite. The actual length (AL) and electronic length (EL) measurements were made on each specimen separately with both devices with the aid of a k-type file. RESULTS: The results obtained with each EAL were compared with the corresponding actual length. The statistical analysis of the results showed EAL reliability to prevent overestimated working length to be 100% for the Mini Apex Locator and 97.44% for the Root ZX-II, within a tolerance of +/-0.5 mm into account. A paired sample t test showed that there was no statistically significant difference between the accuracy of the devices (P = .5841). CONCLUSION: The results of this study indicate that the Root ZX-II and Mini Apex Locator are accurate devices to prevent overestimated working length.


Asunto(s)
Odontometría/instrumentación , Ápice del Diente/anatomía & histología , Cavidad Pulpar/anatomía & histología , Cavidad Pulpar/diagnóstico por imagen , Humanos , Odontometría/métodos , Radiografía , Reproducibilidad de los Resultados , Ápice del Diente/diagnóstico por imagen
2.
J Endod ; 32(6): 560-2, 2006 Jun.
Artículo en Inglés | MEDLINE | ID: mdl-16728251

RESUMEN

The objective of this study was to compare the accuracy of the Root-ZX and Novapex electronic apex locators (EALs) in locating the apical foramen. Forty extracted human teeth were used. The coronal portion of each canal was flared using Gates Glidden drill. Canals were irrigated with 2.5% sodium hypochlorite. The actual length (AL) and electronic length (EL) measurements were made on each specimen separately with both devices with an aid of a K-type file. The results obtained with each EAL were compared with the corresponding control length. The statistical analysis of the results showed EAL reliability in detecting the apical foramen to 89.7% for the Root-ZX and 82.1% for the Novapex, taking the tolerance of +/-0.5 mm into account. A paired sample t test showed that there was no statistically significant difference between the accuracy of the devices (p = 0.4305). The results of this study indicate that the Root-ZX and Novapex are useful and accurate devices for the apical foramen location.


Asunto(s)
Instrumentos Dentales , Cavidad Pulpar/anatomía & histología , Odontometría/instrumentación , Ápice del Diente/anatomía & histología , Humanos
SELECCIÓN DE REFERENCIAS
DETALLE DE LA BÚSQUEDA