Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Más filtros











Intervalo de año de publicación
1.
Artículo en Chino | WPRIM (Pacífico Occidental) | ID: wpr-1017038

RESUMEN

Objective To experimentally verify the precision and accuracy of determining tritiated water in ambient air using the desiccant adsorption sampling–high temperature negative pressure desorption of liquid water (containing HTO)–liquid Scintillation counter method, and to provide technical support for developing standard methods for monitoring tritiated water in ambient air. Methods The relative standard deviation and recovery of multi-group samples were verified by collecting, testing, and analyzing environmental samples with different activity concentrations. The uncertainty of the method was evaluated, the main uncertainty components were identified, and the reliability of measurement results was analyzed. Through experimental comparison of different methods, the differences in the test results of different methods were examined. Results The relative standard deviation of multiple samples ranged from 6.7% to 7.9%, the recovery ranged from 95.7% to 97.3%, and the uncertainty was greatly affected by the sample counting rate, with no significant difference as compared to condensation sampling method. Conclusion The precision and accuracy of this method meet the requirements of environmental authorities for monitoring tritiated water in ambient air, and it can be widely used in the monitoring of tritiated water in ambient air.

SELECCIÓN DE REFERENCIAS
DETALLE DE LA BÚSQUEDA