RESUMEN
INTRODUCTION: Simultaneous mechanical ventilation of several patients with a single ventilator might reduce the deficit of these devices for the care of patients with acute respiratory failure due to Covid-19. OBJECTIVE: To communicate the results of a mechanical ventilation exercise with a ventilator in a lung simulator, and simultaneously in two and four. RESULTS: No statistically significant differences were observed between programmed, recorded and measured positive end-expiratory pressure, mean airway pressure and peak pressure, except when simultaneously ventilating four lung simulators. CONCLUSIONS: Simultaneous mechanical ventilation should be implemented by medical personnel with experience in the procedure, be restricted to two patients and carried out in the intensive care unit.
INTRODUCCIÓN: La ventilación mecánica simultánea a varios pacientes con un solo ventilador podría disminuir el déficit de esos dispositivos para atender a los enfermos con insuficiencia respiratoria aguda por Covid-19. OBJETIVO: Comunicar los resultados de un ejercicio de ventilación mecánica con un ventilador en un simulador de pulmón, y simultáneamente en dos y cuatro. RESULTADOS: No se observaron diferencias estadísticamente significativas entre la presión positiva al final de la espiración, presión media de la vía aérea y presión pico programadas, registradas y medidas, excepto al ventilar simultáneamente cuatro simuladores de pulmón. CONCLUSIONES: La ventilación mecánica simultánea debe ser instaurada por personal médico con experiencia en el procedimiento, restringirse a dos pacientes y ser realizada en la unidad de cuidados intensivos.