Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Más filtros











Base de datos
Intervalo de año de publicación
1.
J Appl Clin Med Phys ; 9(3): 1-15, 2008 Jun 23.
Artículo en Inglés | MEDLINE | ID: mdl-18716583

RESUMEN

The electron Monte Carlo (eMC) dose calculation algorithm of the Eclipse treatment planning system is based heavily upon Monte Carlo simulation of the linac head and modeling of the linac beam characteristics with minimal measurement of beam data. Commissioning of the eMC algorithm on multiple identical linacs provided a unique opportunity to systematically evaluate the algorithm with actual measurements of clinically relevant beam and dose parameters. In this study, measured and eMC calculated dose distributions were compared both along and perpendicular to electron beam direction for electron energy/applicator/depth combination using measurement data from four Varian 21EX CLINAC linear accelerator (Varian Medical System, Palo Alto, CA). Cutout factors for sizes down to 3 x 3 cm were also compared. Comparisons between the measurement and the eMC calculated values show that the R90, R80, R50, and R10 values mostly agree within 3 mm. Measure and Calculated bremsstrahlung dose Dx correlates well statistically although eMC calculated Dx values are consistently smaller than the measured, with maximum discrepancy of 1% for the 20 MeV electron beams. Surface dose agrees mostly within 2%. Field width and penumbra agree mostly within 3mm. Calculation grid size is found to have a significant effect on the dose calculation. A grid size of 5 mm can produce erroneous dose distributions. Using a grid size of 2.5 mm and a 3% accuracy specified for the eMC to stop calculation iteration, the absolute output agrees with measurements within 3% for field sizes of 5 x 5 cm or larger. For cutout of 3 x 3 cm, however, the output disagreement can reach 8%. Our result indicate that eMC algorithm in Eclipse provides acceptable agreement with measurement data for most clinical situations. Calculation grid size of 2.5 mm or smaller is recommended.


Asunto(s)
Algoritmos , Electrones/uso terapéutico , Método de Montecarlo , Aceleradores de Partículas , Planificación de la Radioterapia Asistida por Computador/métodos
SELECCIÓN DE REFERENCIAS
DETALLE DE LA BÚSQUEDA