1.
Pacing Clin Electrophysiol ; 45(5): 666-673, 2022 05.
Artículo en Inglés | MEDLINE | ID: mdl-35417608

RESUMEN

AIMS: Conduction system pacing has gained steady interest over recent years. While the majority of tools and delivery techniques were developed for His bundle pacing (HBP), the feasibility and reproducibility of using these similar tools for left bundle branch pacing (LBBP) has yet to be determined. We describe our technique for performing LBBP using the Abbott Agilis HisPro™ Steerable Catheter. METHODS AND RESULTS: A series of 22 patients with a mean age of 71.7 years (16 males, 72.7%), underwent LBBP procedure with this catheter between May and October 2021. Nineteen patients (86%) had successful LBBP lead implantation. There were no major complications or mortality. CONCLUSION: The Agilis HisPro™ catheter along with the stylet driven Tendril STS Model 2088TC lead is a safe and feasible delivery system for LBBP.


Asunto(s)
Fascículo Atrioventricular , Terapia de Resincronización Cardíaca , Anciano , Estimulación Cardíaca Artificial/métodos , Terapia de Resincronización Cardíaca/métodos , Catéteres , Electrocardiografía , Sistema de Conducción Cardíaco , Humanos , Masculino , Reproducibilidad de los Resultados