1.
J Interv Card Electrophysiol ; 52(1): 1-8, 2018 Jun.
Artículo en Inglés | MEDLINE | ID: mdl-29460232

RESUMEN

PURPOSE: The optimal radiofrequency (RF) power and lesion duration using contact force (CF) sensing catheters for atrial fibrillation (AF) ablation are unknown. We evaluate 50 W RF power for very short durations using CF sensing catheters during AF ablation. METHODS: We evaluated 51 patients with paroxysmal (n = 20) or persistent (n = 31) AF undergoing initial RF ablation. RESULTS: A total of 3961 50 W RF lesions were given (average 77.6 ± 19.1/patient) for an average duration of only 11.2 ± 3.7 s. As CF increased from < 10 to > 40 g, the RF application duration decreased from 13.7 ± 4.4 to 8.6 ± 2.5 s (p < 0.0005). Impedance drops occurred in all ablations, and for patients in sinus rhythm, there was loss of pacing capture during RF delivery suggesting lesion creation. Only 3% of the ablation lesions were at < 5 g and 1% at > 40 g of force. As CF increased, the force time integral (FTI) increased from 47 ± 24 to 376 ± 102 gs (p < 0.0005) and the lesion index (LSI) increased from 4.10 ± 0.51 to 7.63 ± 0.50 (p < 0.0005). Both procedure time (101 ± 19.7 min) and total RF energy time (895 ± 258 s) were very short. For paroxysmal AF, the single procedure freedom from AF was 86% at 1 and 2 years. For persistent AF, it was 83% at 1 year and 72% at 2 years. There were no complications. CONCLUSIONS: Short duration 50 W ablations using CF sensing catheters are safe and result in excellent long-term freedom from AF for both paroxysmal and persistent AF with short procedure times and small amounts of total RF energy delivery.


Asunto(s)
Fibrilación Atrial/cirugía , Catéteres Cardíacos , Ablación por Catéter/instrumentación , Tempo Operativo , Venas Pulmonares/cirugía , Anciano , Fibrilación Atrial/diagnóstico por imagen , Ablación por Catéter/métodos , Estudios de Cohortes , Electrocardiografía/métodos , Diseño de Equipo , Estudios de Factibilidad , Femenino , Estudios de Seguimiento , Humanos , Masculino , Persona de Mediana Edad , Monitoreo Intraoperatorio/métodos , Resultado del Tratamiento
SELECCIÓN DE REFERENCIAS
DETALLE DE LA BÚSQUEDA