J Biomech ; 38(10): 2113-7, 2005 Oct.
Artículo en Inglés | MEDLINE | ID: mdl-16084212

RESUMEN

An electromagnetic position tracking device was evaluated to determine its static and dynamic accuracy and reliability for applications related to measuring in vivo joint kinematics. The device detected the position and orientation of small coiled sensors, maintained in an electromagnetic field. System output was measured against known translations or rotations throughout the measurement volume. Average translational errors during static testing were 0.1 +/- 0.04, 0.2 +/- 0.17, and 0.8 +/- 0.81 mm (mean+/-SD) for sensors 50, 300, and 550 mm away from the field generator, respectively. Average rotational errors were 0.4 +/- 0.31 degrees, 0.4 +/- 0.21 degrees, and 0.9 +/- 0.85 degrees (mean +/- SD) for sensors located at the same distances. Since we intended to use this system in an animal walking on a treadmill, we incrementally moved the sensors under various treadmill conditions. The effects of treadmill operation on translational accuracy were found to be negligible. The effects of dynamic motions on sensor-to-sensor distance were also assessed for future data collection in the animal. Sensor-to-sensor distance showed standard deviations of 2.6 mm and a range of 13 mm for the highest frequency tested (0.23 Hz). We conclude that this system is useful for static or slow dynamic motions, but is of limited use for obtaining gait kinematics at higher speeds.


Asunto(s)
Fenómenos Electromagnéticos/instrumentación , Articulaciones/fisiología , Fenómenos Biomecánicos , Marcha , Humanos
SELECCIÓN DE REFERENCIAS
DETALLE DE LA BÚSQUEDA