1.
Clin Oral Implants Res ; 24(7): 758-62, 2013 Jul.
Artículo en Inglés | MEDLINE | ID: mdl-22568699

RESUMEN

OBJECTIVE: The objective of this clinical study was to evaluate survival and incidence of complications for metal-ceramic and all-ceramic implant-supported fixed dental prostheses (FDPs) and tooth-implant-supported FDPs. MATERIAL AND METHODS: One-hundred and sixty-six FDPs placed in 132 patients from a prospective study were selected for this retrospective analysis. Included were 91 conventional implant-supported FDPs with implant support on both extremities, 27 implant-supported cantilever FDPs, and 48 tooth-implant-supported FDPs. All restorations were veneered with ceramic. Twenty-six FDPs had a zirconia framework and 140 had a metal framework. Kaplan-Meier analysis was performed to estimate FDP success defined as complication-free survival and the Cox regression model was used to isolate risk factors for the most frequent complications. RESULTS: Within a median follow-up of 1 year and 2 months, three failures were caused by a failed implant (n = 2) and by extended chipping of the veneer (n = 1). In contrast with this low incidence of failure was a high incidence of complications including chipping (n = 29), loss of retention (n = 35), and abutment fractures (n = 2). Multivariate survival analysis revealed a significantly greater incidence of chipping for males and a tendency to increased incidence of chipping for zirconia-based FDPs. The incidence of loss of retention tended to be less for tooth-implant-supported FDPs, for which semi-permanent cement was the only significant risk factor, with a hazard ratio of almost 5. CONCLUSIONS: As chipping of the ceramic veneer was the most frequent complication leading to substantial aftercare, improvements of ceramic veneers are desirable for zirconia-based and metal-based FDPs.


Asunto(s)
Implantes Dentales/efectos adversos , Prótesis Dental de Soporte Implantado/efectos adversos , Dentadura Parcial Fija/efectos adversos , Diente/fisiología , Adulto , Anciano , Anciano de 80 o más Años , Cerámica/química , Cementos Dentales/química , Materiales Dentales/química , Fracaso de la Restauración Dental/estadística & datos numéricos , Coronas con Frente Estético , Diseño de Dentadura , Retención de Dentadura/estadística & datos numéricos , Femenino , Estudios de Seguimiento , Humanos , Masculino , Aleaciones de Cerámica y Metal/química , Persona de Mediana Edad , Estudios Prospectivos , Estudios Retrospectivos , Propiedades de Superficie , Análisis de Supervivencia , Adulto Joven , Circonio/química
SELECCIÓN DE REFERENCIAS
DETALLE DE LA BÚSQUEDA