1.
Med Phys ; 34(7): 2801-6, 2007 Jul.
Artículo en Inglés | MEDLINE | ID: mdl-17821987

RESUMEN

A new 125I source under the name Braquibac has been developed in Argentina for interstitial brachytherapy applications. The aim of this work is to study the new seed's design and to calculate its dosimetric parameters. Radiographic and destructive tests were carried out on inactive seeds to determine the physical characteristics of the source. Values of g(r), Lambda, F(r, Theta), and phi(an)(r), were obtained in water and air by simulation using the MCNP5 Monte Carlo code according to the methodology recommended in TG-43 and updated in TG-43U1. The dose rate constant was determined to be 0.937+/-0.004 cGy h(-1) U(-1) (overall statistical uncertainty +/-2.7%). Sk per unity activity was calculated to be 0.671+/-0.003 cGy cm2 h(-1) mCi(-1) by simulation of the seed in dry air using point detectors. Spectroscopic studies for both the new and the Amersham model 6711 seed were performed using an HPGe planar detector. The emission spectra of both seeds proved to be very similar. The anisotropy of the total photon intensity in air was measured in planes containing the seed's short and long axes using the HPGe detector. The minimum photon intensity for the new seed was 31.14+/-3.10% of the transverse intensity.


Asunto(s)
Braquiterapia , Radioisótopos de Yodo , Anisotropía , Radioisótopos de Yodo/uso terapéutico , Método de Montecarlo , Radiometría
SELECCIÓN DE REFERENCIAS
DETALLE DE LA BÚSQUEDA