Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 5 de 5
Filtrar
Más filtros











Intervalo de año de publicación
1.
Arch Cardiol Mex ; 90(2): 142-147, 2020.
Artículo en Inglés | MEDLINE | ID: mdl-32897265

RESUMEN

In this paper, we describe our coronary stent (INC-1) design and development, the way that we found the specific characteristics needed for our device including biophysics aspects, design, finite element testing, manufacturing, and mechanic trials, we submitted and animal models experiences. The stent platform was cobalt-chromium L605 (Co-Cr), with struts thickness of 80 µm, balloon expandable. We placed the coronary stent INC-1 on a rabbit and dog models so we can evaluate efficacy and security of the device in relationship to its biomechanical properties including navigation capacity, traceability, symmetric expansion, and safety, as well as endothelial attachment, thrombogenicity, and lack of involvement of secondary branches in these models. We succeeded in efficacy and safety of the device after fluoroscopy-guided implant proving excellent capacity of release system, traceability, fluoroscopic visualization, symmetric expansion, and complete endothelial attach. Furthermore, we obtained a good post-implant balloon withdrawal, functional integrity, and no vascular complications. We observed adequate clinical evolution 3 weeks after the stent implantation.


En esta publicación se describen el diseño y el desarrollo de un stent coronario (INC-1), la forma en que se determinaron las características biofísicas deseadas, el diseño y las pruebas computacionales por elemento finito, su fabricación, así como las pruebas mecánicas a las que se sometió el prototipo y por último las primeras experiencias en modelos animales. La plataforma del stent desarrollado fue de cobalto y cromo L605 (Co-Cr), con struts de 80 µm de grosor, expandido por balón. La colocación del stent coronario INC-1 se realizó en un modelo de conejo y uno canino con el objetivo de evaluar el éxito técnico y de seguridad del stent en relación con las propiedades biomecánicas (capacidad de navegación, maniobrabilidad, expansión simétrica del stent) y de seguridad (fijación a la pared vascular y ausencia de compromiso de ramas secundarias y formación de trombos en el momento de la colocación del stent en el modelo animal). Se obtuvo éxito técnico y de seguridad del stent después del implante guiado por fluoroscopia, lo que demostró adecuada capacidad del sistema liberador, excelente maniobrabilidad, adecuada visualización fluoroscópica, expansión simétrica y fijación completa a la pared vascular, capacidad para retirar el balón desinflado después del implante, e integridad funcional y sin complicaciones vasculares. La evolución clínica fue favorable a tres semanas del seguimiento planeado.


Asunto(s)
Vasos Coronarios/cirugía , Diseño de Prótesis , Stents , Animales , Aleaciones de Cromo/química , Enfermedad de la Arteria Coronaria/cirugía , Perros , Fluoroscopía , Humanos , Modelos Animales , Conejos
2.
Arch. cardiol. Méx ; 90(2): 142-147, Apr.-Jun. 2020. graf
Artículo en Inglés | LILACS | ID: biblio-1131023

RESUMEN

Abstract In this paper, we describe our coronary stent (INC-1) design and development, the way that we found the specific characteristics needed for our device including biophysics aspects, design, finite element testing, manufacturing, and mechanic trials, we submitted and animal models experiences. The stent platform was cobalt-chromium L605 (Co-Cr), with struts thickness of 80 μm, balloon expandable. We placed the coronary stent INC-1 on a rabbit and dog models so we can evaluate efficacy and security of the device in relationship to its biomechanical properties including navigation capacity, traceability, symmetric expansion, and safety, as well as endothelial attachment, thrombogenicity, and lack of involvement of secondary branches in these models. We succeeded in efficacy and safety of the device after fluoroscopy-guided implant proving excellent capacity of release system, traceability, fluoroscopic visualization, symmetric expansion, and complete endothelial attach. Furthermore, we obtained a good post-implant balloon withdrawal, functional integrity, and no vascular complications. We observed adequate clinical evolution 3 weeks after the stent implantation.


Resumen En esta publicación se describen el diseño y el desarrollo de un stent coronario (INC-1), la forma en que se determinaron las características biofísicas deseadas, el diseño y las pruebas computacionales por elemento finito, su fabricación, así como las pruebas mecánicas a las que se sometió el prototipo y por último las primeras experiencias en modelos animales. La plataforma del stent desarrollado fue de cobalto y cromo L605 (Co-Cr), con struts de 80 μm de grosor, expandido por balón. La colocación del stent coronario INC-1 se realizó en un modelo de conejo y uno canino con el objetivo de evaluar el éxito técnico y de seguridad del stent en relación con las propiedades biomecánicas (capacidad de navegación, maniobrabilidad, expansión simétrica del stent) y de seguridad (fijación a la pared vascular y ausencia de compromiso de ramas secundarias y formación de trombos en el momento de la colocación del stent en el modelo animal). Se obtuvo éxito técnico y de seguridad del stent después del implante guiado por fluoroscopia, lo que demostró adecuada capacidad del sistema liberador, excelente maniobrabilidad, adecuada visualización fluoroscópica, expansión simétrica y fijación completa a la pared vascular, capacidad para retirar el balón desinflado después del implante, e integridad funcional y sin complicaciones vasculares. La evolución clínica fue favorable a tres semanas del seguimiento planeado.


Asunto(s)
Humanos , Animales , Perros , Conejos , Diseño de Prótesis , Stents , Vasos Coronarios/cirugía , Enfermedad de la Arteria Coronaria/cirugía , Fluoroscopía , Aleaciones de Cromo/química , Modelos Animales
3.
Arch Cardiol Mex ; 90(2): 163-168, 2020.
Artículo en Inglés | MEDLINE | ID: mdl-32459214

RESUMEN

In this paper we describe our coronary stent (INC-1) design and development, the way that we found the specific characteristics needed for our device including biophysics aspects, design, finite element testing, manufacturing and mechanic trials we submitted and animal models experiences. The stent platform was cobalt-chromium L605 (Co-Cr), with struts thickness of 80 µm, balloon expandable. We placed the coronary stent INC-1 on a rabbit and dog models so we can evaluate efficacy and security of the device in relationship to its biomechanical properties including navigation capacity, traceability, symmetric expansion and safety. As well as endothelial attachment, thrombogenicity and lack of involvement of secondary branches in these models. We succeeded in efficacy and safety of the device after fluoroscopy-guided implant proving excellent capacity of release system, traceability, fluoroscopic visualization, symmetric expansion and complete endothelial attach. Also, we obtained a good post-implant balloon withdrawal, functional integrity and no vascular complications. We observed adequate clinical evolution 3 weeks after the stent implantation.


En esta publicación se describen el diseño y el desarrollo de un stent coronario (INC-1), la forma en que se determinaron las características biofísicas deseadas, el diseño y las pruebas computacionales por elemento finito, su fabricación, así como las pruebas mecánicas a las que se sometió el prototipo y por último las primeras experiencias en modelos animales. La plataforma del stent desarrollado fue de cobalto y cromo L605 (Co-Cr), con struts de 80 µm de grosor, expandido por balón. La colocación del stent coronario INC-1 se realizó en un modelo de conejo y uno canino con el objetivo de evaluar el éxito técnico y de seguridad del stent en relación con las propiedades biomecánicas (capacidad de navegación, maniobrabilidad, expansión simétrica del stent) y de seguridad (fijación a la pared vascular y ausencia de compromiso de ramas secundarias y formación de trombos en el momento de la colocación del stent en el modelo animal). Se obtuvo éxito técnico y de seguridad del stent después del implante guiado por fluoroscopia, lo que demostró adecuada capacidad del sistema liberador, excelente maniobrabilidad, adecuada visualización fluoroscópica, expansión simétrica y fijación completa a la pared vascular, capacidad para retirar el balón desinflado después del implante, e integridad funcional y sin complicaciones vasculares. La evolución clínica fue favorable a tres semanas del seguimiento planeado.


Asunto(s)
Enfermedad de la Arteria Coronaria/cirugía , Diseño de Prótesis , Stents , Animales , Modelos Animales de Enfermedad , Perros , Fluoroscopía , Humanos , Conejos
4.
Rev Invest Clin ; 72(1): 32-36, 2020.
Artículo en Inglés | MEDLINE | ID: mdl-32132738

RESUMEN

BACKGROUND: The cost of performing a percutaneous coronary intervention is considerably high for the patient as well as for health systems, which have promoted the development of local technology to help meet the need for these devices. METHODS: The INC-01 bare-metal stent was developed at the National Institute of Cardiology in Mexico City and was first implanted on porcine models with technical success in 100% of the evaluated parameters. PRESENTATION OF CASES: We present the first three cases of patients with ischemic heart disease, to whom the INC-01 bare-metal stent was implanted. Intracoronary ultrasonography was performed post-stent implantation, showing all the characteristics of implant success during evaluation and clinical follow-up. CONCLUSIONS: Angiography and intracoronary ultrasound were carried out demonstrating that the INC-01 bare-metal stent has physical, biological, and histological characteristics similar to those found in commercial metallic stents.


Asunto(s)
Isquemia Miocárdica/cirugía , Diseño de Prótesis , Stents , Adulto , Anciano , Estudios de Seguimiento , Humanos , Masculino , México , Persona de Mediana Edad , Isquemia Miocárdica/diagnóstico por imagen , Isquemia Miocárdica/fisiopatología , Proyectos Piloto , Resultado del Tratamiento , Ultrasonografía
5.
Rev. invest. clín ; 72(1): 32-36, Jan.-Feb. 2020. graf
Artículo en Inglés | LILACS | ID: biblio-1251832

RESUMEN

ABSTRACT Background: The cost of performing a percutaneous coronary intervention is considerably high for the patient as well as for health systems, which have promoted the development of local technology to help meet the need for these devices. Methods: The INC-01 bare-metal stent was developed at the National Institute of Cardiology in Mexico City and was first implanted on porcine models with technical success in 100% of the evaluated parameters. Presentation of Cases: We present the first three cases of patients with ischemic heart disease, to whom the INC-01 bare-metal stent was implanted. Intracoronary ultrasonography was performed post-stent implantation, showing all the characteristics of implant success during evaluation and clinical follow-up. Conclusions: Angiography and intracoronary ultrasound were carried out demonstrating that the INC-01 bare-metal stent has physical, biological, and histological characteristics similar to those found in commercial metallic stents.


Asunto(s)
Humanos , Masculino , Adulto , Persona de Mediana Edad , Anciano , Diseño de Prótesis , Stents , Isquemia Miocárdica/cirugía , Proyectos Piloto , Estudios de Seguimiento , Ultrasonografía , Resultado del Tratamiento , Isquemia Miocárdica/fisiopatología , Isquemia Miocárdica/diagnóstico por imagen , México
SELECCIÓN DE REFERENCIAS
DETALLE DE LA BÚSQUEDA