Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 2 de 2
Filtrar
Más filtros











Intervalo de año de publicación
1.
Arq Bras Oftalmol ; 79(5): 312-314, 2016.
Artículo en Inglés | MEDLINE | ID: mdl-27982210

RESUMEN

PURPOSE:: To compare central corneal thickness (CCT) measurements of healthy individuals obtained with ultrasonic pachymetry (UP) and non-contact specular microscopy (NCSM). METHOD:: In total, 148 eyes of 74 subjects with no ocular or systemic diseases were included in the study. Central corneal thickness measurements of all patients performed with UP and NCCM were compared. RESULTS:: A total of 74 subjects (38 females) were included in this study. The mean age was 45.2 ± 18.4 (range 12-85) years. The mean central corneal thickness of all 148 eyes was 546.9 ± 40 µm with UP and 510.8 ± 42 µm with NCSM. The mean central corneal thickness measured with NCSM was 35 µm thinner than that measured with UP (p<0.001). A high degree of agreement was found between the two methods (r=0.942, p<0.001). CONCLUSIONS:: Our results suggest that NCSM measures thinner corneas than UP and that the correction formula we identified should be applied when comparing between these two devices.


Asunto(s)
Córnea/anatomía & histología , Paquimetría Corneal/instrumentación , Microscopía/instrumentación , Ultrasonografía/instrumentación , Adolescente , Adulto , Anciano , Anciano de 80 o más Años , Niño , Paquimetría Corneal/métodos , Femenino , Humanos , Modelos Lineales , Masculino , Microscopía/métodos , Persona de Mediana Edad , Valores de Referencia , Reproducibilidad de los Resultados , Estadísticas no Paramétricas , Ultrasonografía/métodos , Adulto Joven
2.
Arq. bras. oftalmol ; Arq. bras. oftalmol;79(5): 312-314, Sept.-Oct. 2016. graf
Artículo en Inglés | LILACS | ID: biblio-827974

RESUMEN

ABSTRACT Purpose: To compare central corneal thickness (CCT) measurements of healthy individuals obtained with ultrasonic pachymetry (UP) and non-contact specular microscopy (NCSM). Method: In total, 148 eyes of 74 subjects with no ocular or systemic diseases were included in the study. Central corneal thickness measurements of all patients performed with UP and NCCM were compared. Results: A total of 74 subjects (38 females) were included in this study. The mean age was 45.2 ± 18.4 (range 12-85) years. The mean central corneal thickness of all 148 eyes was 546.9 ± 40 μm with UP and 510.8 ± 42 μm with NCSM. The mean central corneal thickness measured with NCSM was 35 μm thinner than that measured with UP (p<0.001). A high degree of agreement was found between the two methods (r=0.942, p<0.001). Conclusions: Our results suggest that NCSM measures thinner corneas than UP and that the correction formula we identified should be applied when comparing between these two devices.


RESUMO Objetivo: Comparar as medidas de espessura central corneana (CCT) de indivíduos saudáveis obtidos pela paquimetria ultrassônica (UP) e microscopia especular sem contato (NCSM). Método: Cento e quarenta e oito olhos de 74 indivíduos que não tinham doenças oculares ou sistêmicas foram incluídos no estudo. Medidas da espessura central corneana de todos os pacientes foram comparadas entre UP (SP 100, Tomey, Nagoya, Japão) e do NCSM (NSP-9900, Konan Medical, Inc., Hyogo, Japão). Resultados: Um total de 74 indivíduos (38 mulheres) foram incluídos neste estudo. A idade média foi de 45,2 ± 18,4 (variação 12-85) anos. A medida média da espessura central corneana de todos os 148 olhos foi 546,9 ± 40 μm com UP e 510,8 ± 42 μm com NCSM. A espessura central corneana média avaliada pelo NCSM foi de 35 μm mais fina do que a UP (p<0,001). Foi encontrado um elevado grau de concordância entre os dois métodos (r=0,942, p<0,001). Conclusões: Nossos resultados sugerem que a microscopia especular sem contato mede córneas mais finas em comparação com a UP e que o fator de correção identificado deve ser aplicado ao fazer comparações entre esses dois aparelhos.


Asunto(s)
Humanos , Masculino , Femenino , Niño , Adolescente , Adulto , Persona de Mediana Edad , Anciano , Anciano de 80 o más Años , Adulto Joven , Ultrasonografía/instrumentación , Córnea/anatomía & histología , Paquimetría Corneal/instrumentación , Microscopía/instrumentación , Valores de Referencia , Modelos Lineales , Reproducibilidad de los Resultados , Ultrasonografía/métodos , Estadísticas no Paramétricas , Paquimetría Corneal/métodos , Microscopía/métodos
SELECCIÓN DE REFERENCIAS
DETALLE DE LA BÚSQUEDA