Technol Health Care ; 18(1): 19-29, 2010.
Artículo en Inglés | MEDLINE | ID: mdl-20231800

RESUMEN

The current study evaluated whether 3-D fluoroscopic imaging is capable of adequate visualisation, reduction and effective guidance of implant placement during a minimally-invasive screw-fixation (MISF) procedure in the treatment of proximal phalanx fractures. A comparison with conventional intraoperative 2-D imaging was performed in a cadaveric model. Conventional 2-D and 3-D imaging series were performed following the creation of proximal phalanx fractures, reduction and fixation, pre- and post-operatively. For both imaging modalities, attention was paid to A) correct reduction, B) screw-placement and, if present, C) intra-articular offset of fracture edges.The results revealed no related overall advantage of the 3-D imaging system over normal 2-D fluoroscopy at the proximal phalanx region. The authors conclude that, given the cost of 3-D imaging technology, as well as the increased time required for image acquisition, its routine use in the treatment of phalanx fracture cases is not justifiable at present.


Asunto(s)
Tornillos Óseos , Traumatismos de los Dedos/diagnóstico por imagen , Fracturas Óseas/diagnóstico por imagen , Imagenología Tridimensional/instrumentación , Cadáver , Fluoroscopía/métodos , Humanos , Procesamiento de Imagen Asistido por Computador , Resultado del Tratamiento , Ultrasonografía
SELECCIÓN DE REFERENCIAS