1.
Appl Radiat Isot ; 70(9): 2232-6, 2012 Sep.
Artículo en Inglés | MEDLINE | ID: mdl-22429557

RESUMEN

The response of a Vacutec 70129 ionization chamber was calculated using the PENELOPE-2008 Monte Carlo code and compared to experimental data. The filling gas mixture composition and its pressure have been determined using IC simulated response adjustment to experimental results. The Monte Carlo simulation revealed a physical effect in the detector response to photons due to the presence of xenon in the chamber. A very good agreement is found between calculated and experimental calibration coefficients for 17 radionuclides.


Asunto(s)
Modelos Estadísticos , Radiometría/instrumentación , Radiometría/normas , Calibración , Simulación por Computador , Análisis de Falla de Equipo , Internacionalidad , Dosis de Radiación , Estándares de Referencia , Valores de Referencia
2.
Appl Radiat Isot ; 66(6-7): 780-5, 2008.
Artículo en Inglés | MEDLINE | ID: mdl-18397830

RESUMEN

This work relates to the study and characterization of the response function of an X-ray spectrometry system. The intrinsic efficiency of a Si(Li) detector has been simulated with the Monte Carlo codes MCNP and GEANT4 in the photon energy range of 2.6-59.5keV. After finding it necessary to take a radiograph of the detector inside its cryostat to learn the correct dimensions, agreement within 10% between the simulations and experimental measurements with several point-like sources and PIXE results was obtained.