RESUMEN
Comparative analysis of 1H NMR spectroscopy and refractometry with respect to their application for investigating the distribution of nonelectrolytes of n-alcohol series (ethanol, 1,2-propanediol, glycerol) and polyethylene glycols (PEGs) with molecular masses of 400, 600, 1500 between human erythrocytes and extracellular medium was performed. The distribution coefficients (Q) for solutions of ethanol, 1,2-propanediol, glycerol, PEG-400, PEG-600 and PEG-1500 were obtained. The Q values decreased with the increase in the nonelectrolyte molecular mass from 1.23+/-0.12 for ethanol to 0.40+/-0.08 for PEG-1500 (1H NMR spectroscopy) and from 2.6+/-0.12 for ethanol to 0.23+/-0.03 for PEG-1500 (refractometry). It was shown that 1H-NMR high-resolution spectroscopy ensures more precise determination of Q values for nonelectrolytes with low molecular masses; for PEGs with high molecular masses, the accuracy of Q value calculation by this method was about 20%. On the contrary, refractometry can be used for investigating substances with high molecular masses; the error of Q value determination for solution of low-refractive substances, such as ethanol, may be more than 50%.