1.
Nihon Hoshasen Gijutsu Gakkai Zasshi ; 63(11): 1246-52, 2007 Nov 20.
Artículo en Japonés | MEDLINE | ID: mdl-18187909

RESUMEN

A variety of radiation hazards resulting from interventional radiology (IVR) have been reported in recent years. Particularly affected are the skin and the crystalline lens, with their high radiation sensitivity. During neurological interventions, the radiological technologist should consider decreasing radiation exposure. We found exposure projections where the exposure dose became a radiation hazard for the crystalline lens, and examined an efficient method of cover for the exposure projections used for neurological interventions. The exposure projection for maximum crystalline lens radiation exposure was a lateral projection. In the crystalline lens the maximum exposure to radiation was on the X-ray tube side. The method of defense adopted was that of installing a lead plate of the appropriate shape on the surface of the X-ray tube collimator. In other exposure projections, this cover did not become a redundant shadow. With the cover that was created, the X-ray side crystalline lens lateral projection could be defended effectively.


Asunto(s)
Arterias Cerebrales/diagnóstico por imagen , Cristalino/efectos de la radiación , Protección Radiológica/instrumentación , Radiología Intervencionista , Humanos , Radiografía
SELECCIÓN DE REFERENCIAS