Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Más filtros











Base de datos
Intervalo de año de publicación
1.
Mol Diagn Ther ; 2024 Aug 02.
Artículo en Inglés | MEDLINE | ID: mdl-39093547

RESUMEN

BACKGROUND: Next-generation sequencing is widely used for comprehensive molecular profiling for many cancers including lung cancer. However, the complex workflows and long turnaround times limit its access and utility. ChromaCode's High Definition PCR Non-Small Cell Lung Cancer Panel (HDPCR™ NSCLC Panel) is a low-cost, rapid turnaround, digital polymerase chain reaction assay that is designed to detect variants in nine NSCLC genes listed in National Comprehensive Cancer Network guidelines. METHODS: This assay uses TaqMan® probe limiting chemistry and proprietary analysis software to enable multi-target detection within a single-color channel. We compared the performance of the HDPCR™ NSCLC Panel against an in-house, laboratory-developed, targeted next-generation sequencing panel used in the Molecular Diagnostics Laboratory at the University of Minnesota Medical Center to detect biomarkers for NSCLC. RESULTS: The overall accuracy of the HDPCR panel was 99.48% (95% confidence interval 99.01-99.76) with a sensitivity of 95.35% (95% confidence interval 88.52-98.72) and a specificity of 99.69% (95% confidence interval 99.29-99.90). The HDPCR wet lab workflow was 4 h, and the time to generate variant calls from raw data using the ChromaCode Cloud was 2 minutes. CONCLUSIONS: We demonstrated that the HDPCR™ NSCLC Panel provides timely, comprehensive, and sensitive mutation detection in NSCLC samples with results in less than 24 h.

SELECCIÓN DE REFERENCIAS
DETALLE DE LA BÚSQUEDA