Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Más filtros











Base de datos
Intervalo de año de publicación
1.
JPEN J Parenter Enteral Nutr ; 41(8): 1272-1277, 2017 11.
Artículo en Inglés | MEDLINE | ID: mdl-27488830

RESUMEN

BACKGROUND AND AIMS: The 2 currently available indirect calorimeters, CCM Express Indirect Calorimeter (MedGraphics, St Paul, MN) and Quark RMR ICU Indirect Calorimeter (COSMED, Rome, Italy), have not been validated against a gold standard in mechanically ventilated patients. Our aim was to do so using a gold-standard, modified Tissot bell-spirometer method in mechanically ventilated patients who were hemodynamically, respiratory, and metabolically stable. METHODS: We studied 30 patients undergoing general anesthesia and major gynecological surgery. We measured oxygen consumption ([Formula: see text]O2) and resting energy expenditure (REE) in a randomized, sequential, crossover design with double determination of each device. RESULTS: Compared with the modified Tissot bell-spirometer, the CCM Express Indirect Calorimeter demonstrated a mean Δ-REE of +361 kcal/d, corresponding to a 31% overestimation of energy requirements. Bland-Altman analysis for REE showed a mean (SD) bias of 384 (124) with limits of agreement 142-627 kcal/d. QUARK RMR ICU demonstrated a mean Δ-REE of 81 kcal/d, corresponding to a 7% overestimation of energy requirements. Bland-Altman plot analysis showed a mean (SD) bias of 77 (167) with limits of agreement -249 to 404 kcal/d. CONCLUSIONS: The QUARK RMR ICU Indirect Calorimeter compared better with the gold standard for values of [Formula: see text]O2 and REE than did the CCM Express Indirect Calorimeter in mechanically ventilated patients who were circulatory and respiratory stable. Both indirect calorimeters had low precision.


Asunto(s)
Calorimetría Indirecta/instrumentación , Respiración Artificial , Adulto , Anciano , Metabolismo Basal , Estudios Cruzados , Femenino , Humanos , Unidades de Cuidados Intensivos , Italia , Masculino , Persona de Mediana Edad , Necesidades Nutricionales , Consumo de Oxígeno , Estudios Prospectivos , Reproducibilidad de los Resultados , Tamaño de la Muestra
SELECCIÓN DE REFERENCIAS
DETALLE DE LA BÚSQUEDA