Int J Radiat Oncol Biol Phys ; 18(5): 1215-21, 1990 May.
Artículo en Inglés | MEDLINE | ID: mdl-2112121

RESUMEN

The leakage radiation characteristics of a dedicated intraoperative radiotherapy linear accelerator have been measured on a machine designed to minimize the shielding required to allow it to be placed in an operating room suite. The scattering foil design was optimized to produce a flat beam for the field sizes employed while generating minimal bremsstrahlung contamination over the available energy range. More lead shielding was used in the treatment head than is used in conventional accelerators. A small amount of borated polyethylene shielding was also employed since neutron production was present at measurable levels. The room shielding installed in the operating room was demonstrated to be adequate to treat at least 20 patients each month to an average dose of 20 Gy. The worst case exposure was found to be 73% maximum permissible exposure. Administrative control was required for adjoining areas when calibrations and maintenance were performed.


Asunto(s)
Quirófanos , Aceleradores de Partículas , Protección Radiológica , Radioterapia de Alta Energía , Periodo Intraoperatorio , Neutrones , Protección Radiológica/instrumentación , Radioterapia de Alta Energía/instrumentación , Rayos X