1.
Crit Care Resusc ; 2(1): 92-3, 2000 Mar.
Artículo en Inglés | MEDLINE | ID: mdl-16597291
2.
Anaesth Intensive Care ; 24(4): 423-9, 1996 Aug.
Artículo en Inglés | MEDLINE | ID: mdl-8862637

RESUMEN

There is currently no universally accepted method to monitor circuit function or guidelines for circuit replacement during continuous renal replacement therapies (CRRT). The objectives of this study were to diagnose the causes of circuit failure, identify factors responsible for circuit clotting and determine a predictive monitor of circuit function. The CRRT technique used in this study was continuous venovenous haemodialysis (CVVHD). Continuous monitoring of circuit pressures (pre- and post-haemofilter and their difference: the transfilter pressure gradient) was used to diagnose the causes of circuit failure. In circuits ceasing due to clotting, the factors thought to contribute, anticoagulation, haematocrit and platelet count, were measured at the commencement of CVVHD and every eight hours thereafter until circuit failure. Monitors of circuit function, creatinine clearance and plasma to diafiltrate urea ratio were measured every eight hours and compared to the transfilter pressure gradient. During a three-month period data was collected on five consecutive patients (41 consecutive haemofilters). Clotting of the haemofilter (63%) and air detection chamber (7.5%) were the most common identifiable causes of circuit failure. The duration of their circuit life was described using multiple regression analysis, i.e. hours of filter life = -82.8 + (delta platelet count x 0.25) + (delta haematocrit x 3.6) + (circuit flow [ml/min] x 4) R2 = 0.77. A rise in transfilter pressure gradient and a fall in haemofilter function discriminated clotted filters with falling function (decrease in creatinine clearance and urea ratio) from unclotted filters. In any circuit an increase of 26 mmHg or more in the transfilter pressure gradient accurately predicted circuit failure due to clotting and imminent cessation of function. Increases in platelet count, haematocrit, and low circuit flows are important determinants of haemofilter life. The measurement of transfilter pressure gradient across the haemofilter is an accurate bedside monitor of circuit function.


Asunto(s)
Lesión Renal Aguda/terapia , Cuidados Críticos/métodos , Diálisis Renal/instrumentación , Falla de Equipo , Hematócrito , Humanos , Presión , Estudios Prospectivos
3.
Crit Care Med ; 20(12): 1688-91, 1992 Dec.
Artículo en Inglés | MEDLINE | ID: mdl-1458947

RESUMEN

OBJECTIVES: To determine the interobserver reliability of residents and nurses collecting Acute Physiology and Chronic Health Evaluation (APACHE II) data and the subsequent effect of these data collections on individual patient mortality prediction. DESIGN: In a prospective study, residents and nurses independently collected data to derive APACHE II scores. When their scores differed, a standard score was determined by one of the investigators. SETTING: A general medical and surgical ICU. PATIENTS: A total of 120 consecutive patients were included; of these patients, 79 had standard scores determined because resident and nurse scores differed. MAIN RESULTS: There was overall agreement between the residents and nurses with no significant difference between mean APACHE II scores or mean predicted mortality rates. Intraclass correlation coefficients confirmed good overall agreement between observer groups for predicted mortality rate: resident vs. nurse r2 = .94, resident vs. standard r2 = .94, and nurse vs. standard r2 = .90. However, clinically significant lack of agreement was demonstrated in 5% of the patients by the 95% confidence limits of agreement: resident vs. nurse -14 to +14%, resident vs. standard -10 to +14%, and nurse vs. standard -14 to +20%. CONCLUSIONS: While interobserver variability between resident and nurse data collection has minimal effect on derived predicted mortality rate with large patient groups, significant variability may occur in individual patients. Residents were more accurate data collectors than nurses.


Asunto(s)
Recolección de Datos/normas , Internado y Residencia , Enfermeras y Enfermeros , Índice de Severidad de la Enfermedad , Recolección de Datos/métodos , Humanos , Unidades de Cuidados Intensivos , Mortalidad , Variaciones Dependientes del Observador , Valor Predictivo de las Pruebas , Estudios Prospectivos
SELECCIÓN DE REFERENCIAS
DETALLE DE LA BÚSQUEDA