1.
Radiat Prot Dosimetry ; 180(1-4): 37-41, 2018 Aug 01.
Artículo en Inglés | MEDLINE | ID: mdl-29788424

RESUMEN

A long counter detector was manufactured by the Institute of Advanced Studies (IEAV) and was characterised in the neutron low scattering room at Brazilian National Ionising Radiation Metrology Laboratory (LNMRI/IRD) to deploy a secondary Standard for neutron fluence. The effective centre was measured experimentally with 252Cf+D2O, 252Cf, 241AmBe and 238PuBe neutron sources, having average energies from 0.55 to 4.16 MeV. The experimental arrangement and detector construction were carefully reproduced in Monte Carlo simulations, and the computational results were found to be in good agreement with those from experiment.


Asunto(s)
Americio/normas , Berilio/normas , Californio/normas , Laboratorios/normas , Neutrones , Plutonio/normas , Monitoreo de Radiación/normas , Protección Radiológica/normas , Americio/análisis , Berilio/análisis , Calibración , Californio/análisis , Método de Montecarlo , Plutonio/análisis , Dosis de Radiación
2.
Radiat Prot Dosimetry ; 180(1-4): 56-61, 2018 Aug 01.
Artículo en Inglés | MEDLINE | ID: mdl-29474641

RESUMEN

The standard thermal neutron flux unit, TNF2, in the Brazilian National Ionizing Radiation Metrology Laboratory was rebuilt. Fluence is still achieved by moderating of four 241Am-Be sources with 0.6 TBq each. The facility was again simulated and redesigned with graphite core and paraffin added graphite blocks surrounding it. Simulations using the MCNPX code on different geometric arrangements of moderator materials and neutron sources were performed. The resulting neutron fluence quality in terms of intensity, spectrum and cadmium ratio was evaluated. After this step, the system was assembled based on the results obtained from the simulations and measurements were performed with equipment existing in LNMRI/IRD and by simulated equipment. This work focuses on the characterization of a central chamber point and external points around the TNF2 in terms of neutron spectrum, fluence and ambient dose equivalent, H*(10). This system was validated with spectra measurements, fluence and H*(10) to ensure traceability.


Asunto(s)
Americio/normas , Berilio/normas , Laboratorios/normas , Neutrones , Monitoreo de Radiación/normas , Protección Radiológica/normas , Americio/análisis , Berilio/análisis , Calibración , Método de Montecarlo , Dosis de Radiación
3.
Radiat Prot Dosimetry ; 161(1-4): 185-9, 2014 Oct.
Artículo en Inglés | MEDLINE | ID: mdl-24625545

RESUMEN

The Brazilian Metrology Laboratory of Ionizing Radiations (LNMRI) standard thermal neutron flux facility was designed to provide uniform neutron fluence for calibration of small neutron detectors and individual dosemeters. This fluence is obtained by neutron moderation from four (241)Am-Be sources, each with 596 GBq, in a facility built with blocks of graphite/paraffin compound and high-purity carbon graphite. This study was carried out in two steps. In the first step, simulations using the MCNPX code on different geometric arrangements of moderator materials and neutron sources were performed. The quality of the resulting neutron fluence in terms of spectrum, cadmium ratio and gamma-neutron ratio was evaluated. In the second step, the system was assembled based on the results obtained on the simulations, and new measurements are being made. These measurements will validate the system, and other intercomparisons will ensure traceability to the International System of Units.


Asunto(s)
Protección Radiológica/métodos , Radiometría/instrumentación , Radiometría/métodos , Americio , Berilio , Brasil , Calibración , Carbono/química , Simulación por Computador , Diseño de Equipo , Grafito/química , Método de Montecarlo , Neutrones , Dosis de Radiación , Radiación Ionizante
SELECCIÓN DE REFERENCIAS
DETALLE DE LA BÚSQUEDA