Radiat Prot Dosimetry ; 156(4): 506-13, 2013 Oct.
Artículo en Inglés | MEDLINE | ID: mdl-23567198

RESUMEN

Analyses of the results of calibration of survey meters carried out at the Secondary Standards Dosimetry Laboratory (SSDL) in Ghana over a period of 4 y (2008-2011) are reported. The calibration factors (CFs) of the set of survey meters indicated that ∼91.04 % were within the acceptable limit of ± 20.0 %. A higher percentage of the survey meters indicated CFs in the range of 0.95-1.15 except a few of them which indicated values <0.55. Some of the survey meters also recorded CFs >1.15. The degree of uncertainty in the measurements ranged from 0.03 to 17 % with the majority of them ranging from 0.03 to 6.0 % and a few of them >6.0 %. The results show that most of the survey meters calibrated were within the requirements of the regulations and may provide data for future development of calibration techniques in the country.


Asunto(s)
Radioisótopos de Cesio/análisis , Monitoreo de Radiación/instrumentación , Monitoreo de Radiación/métodos , Radiometría/métodos , Algoritmos , Calibración , Ghana , Humanos , Rayos Láser , Protección Radiológica/métodos , Reproducibilidad de los Resultados