RESUMEN
Objective:To establish a method for the determination of zolpidemtartrate in human plasma by SPE-HPLC. Methods:The samples were extracted by solid phase with a Hypersil BDS C18 column and determined by an internal standard method. The mobile phase was methanol:acetonitrile:water (45 ∶45 ∶110), the flow rate was 1. 0 ml·min-1, the column temperature was 35℃, and the detection wavelength was 254nm. Results:The linear relationship of zolpidemtartrate was within the range of 0. 511 5-4. 092 0 μg· ml-1 . The relative recovery was greater than 90%. The intra-and inter-day RSDs were all less than 10%. Conclusion:The method is quite simple, quick and accurate with little interference, which is suitable for the clinical determination of zolpidemtartrate.